公司產品

550 Coating Thickness Gauge

Product description:

1.Measurement Range: 06000um 2. Accuracy: [(1%~3%)H+1um] Note:H is thickness reading 3. powers down after 60 seconds of no activity. 1、Power-on Keystroke once;Wait until READY! is displayed on the LCD

與:

我們在這裡幫助:
獲得所需答案的簡單方法。
發送電子郵件現在聊天
1.Measurement Range: 0—6000um
2. Accuracy:± [(1%~3%)H+1um] Note:H is thickness reading
3. powers down after 60 seconds of no activity.

1、Power-on
Keystroke once;Wait until READY! is displayed on the LCD. Now the gauge is ready for measurements.
2、Zeroing
Keystroke twice; when BASE ZERO is displayed, place flat on surface; hold steady. After measurement is obtained, DONE is shown and gauge beeps. Then lift gauge.
Note: Please conduct zeroing process after changing base. Zeroing process could be repeated for better results.
3、Two-Point User Calibration
Keystroke three times, enter into the process of tow-point user calibration; when
USER CAL BASE is shown on the LCD, please conduct a measurement on the base first; when
S1 is shown on the LCD, place Shim S1 on the base, conduct a measurement on shim S1 and lift gauge; when
S2 is shown on the LCD, place Shim S2 on the base, conduct a measurement on shim S2 and lift gauge; When DONE is displayed, it is ready for use.
Note: The two-point user calibration process could be repeated for better results.
4、Unit Change
Keystroke four times, convert from ums to mils or vice versa.
5、Reset
Keystroke five times, gauge resets.



admin@hssdtest.com